Back to Top

1
Hi, Guest!

  LOKASI :  Kota Administrasi Jakarta Timur

BAGIKAN :   

Bagikan :
Kontak Kami  
CV. Mitra Nusantara, Adalah Perusahaan yang bergerak di bidang distributor peralatan Laboratorium monitoring lingkungan dan kesehatan kerja

STAPLEX HIGH VOLUME CASCADE IMPACTORS

Update Terakhir
:
27 / 12 / 2023
Min. Pembelian
:
1 Unit

Harga

CALL
Bagikan
:
Dengan hormat, kami adalah perusahaan swasta nasional yang bergerak dibidang distributor serta supilier peralatan laboratorium Kualitas Lingkungan udara, diantaranya produk brand : Staplex, E Instrument, Kitagawa, Shinagawa,, Nova Gas dan EDC

I. STAPLEX
2. NOVA ANALYTICAL SYSTEMS
3. SHINAGAWA
4. HAZ DUST
5. E INSTRUMENT
6. KITAGAWA



Detail STAPLEX HIGH VOLUME CASCADE IMPACTORS

Staplex® Model 234 (4 stage)
Model 235 (5 stage)
Model 236 (6 stage)

Alat Laboratorium Lingkungan

High Volume Cascade Impactors
for use with Staplex® TFIA Series High Volume Air Samplers

(Note: requires Staplex® SH810 8”x 10” Filter Holder Assembly)

  • Measures airborne particulate size distribution
  • For outdoor and occupational environments
  • Separates respirable particulates


DESCRIPTION:

Staplex® Model 234, 235 and 236 High Volume Cascade Impactors are multi stage cascade impactors that attaché to standard High Volume Air Samplers and fractionate suspended particulates into as many as seven size fraction. For Staplex® Model 235, impactor cut-points are nominally: 7.2, 3.0, 1.5, 0.95 and .049 microns. The 8” x 10” high volume filters (see below for Collection Media) collects all particles smaller than 0.49 microns.

Staplex® High Volume Cascade Impactors have been fully calibrated and are field tested by U.S. PA and other research laboratories. Hundreds of Staplex® proven High Volume Cascade Impactors are operating successfully in the field. The Series 230 Impactors provide a simple, accurate, and yet low-cost means of obtaining the complete particle size distribution and respirable mass fraction of suspended particulates in outdoor or occupational environments. The unique slotted impactor design features the ultimate in precision reproducibility, and sharpness of cut-off, combined with compactness, ease of use, and the lowest internal losses

The nominal flow rate is 40 CFM (1.13 SCMM), which falls in the standard range specified by the U.S. EPA. Particles are collected on small, light-weight collection substrates. Both these features provide large particulate samples and unsurpassed measurement sensitivity. The final stages, stages 5 and 6, provide important data points in the sub-micron range.

APPLICATIONS:

Provides air pollution officials and scientists, health physicists, and industrial hygienists with a low-cost, easy-to-use instrument for measuring the size distribution and respirable mass fraction of suspended particulates in outdoor and occupational Environments Other applications include aerosol research and medical measurements.

PARTICLE SIZE DISTRIBUTION:

Particle Size Cut-offs (Microns) at 50% Collection Efficiency for

Spherical Particles with Unity Mass Density @ 25°C and 760mm Hg:

 

Stage No

40CFM

20 CFM

1 7.2 10.2

2

3.0

4.2

3

1.5

2.1

4

0.95

1.4

5

0.49

0.73

6

---

0.41

Hi Vol Filter